GM8036-II Laser scanning spectrum analyzer
GM8036-II laser scanning spectrum analyzer is a PC controlled passive device test and spectrum analysis system is mainly used for coarse wavelength division multiplexing CWDM components, precision WDM DWDM component test, GFF, Interleaver, WSS, AWG & PLC components, optical amplifier EDFA, DWDM system and other general purpose fiber optic measurement and application. The system can expand multi-channel optical power meter and form a tunable laser light source testing system platform with high wave length precision and fast scanning function.
GM8036-II is equipped with PC testing software, can be in accordance with customer needs, flexible build test templates and create test reports. It is a good test platform for passive devices to be arranged and monitored. The fast scanning function of GM8036 can directly display the spectrum of the product. The built-in tunable light source and power meter can be used singly or in combination, so as to realize the function of the scanning spectrum analyzer.
UC INSTRUMENTS’ GM8036-II laser sweep optical spectrum analyzer is a PC control passive components test and spectrum analyze system and offer superior performance for the test of CWDM / DWDM components, optical amplifiers, DWDM system and other general purpose of fiber optical test and measurement applications. The system can be extended multi channel optical power meter,provides a high wavelength accuracy, quick scan function tunable laser source test system platform.
Its flexible PC base application test software can easy build up customers request testing template and create test report. It is also very good testing platform for passive components in process alignment and monitor. Its fast sweep function can direct show products optical spectrum. Build-in tunable laser source, power meter can be use as combine sweep optical spectrum analyze function, and also can function independent.
Specification
Model |
GM8036 |
Tunable Laser Source |
|
Wavelength range |
1527.00 to 1568.00 nm |
Output Power |
>= 5mW |
Wavelength resolution |
1.0 pm/0.01pm |
Absolute wavelength accuracy |
+/- 10 pm, tpy. < 5 pm |
Relative wavelength accuracy |
+/- 5 pm, Typ. +/- 2 pm |
Wavelength repeatability |
+/- 3 pm, typ. +/- 1 pm |
Wavelength stability |
<= +/- 5 pm (constant temperature24 hours, 24 hrs at constant temperature) |
Tuning speed |
<= 2 ms per step |
Power stability |
< +/- 0.1 dB, 24 hours. |
Power repeatability |
+/- 0.05 dB |
Power linearity |
+/- 0.3 dB |
Power flatness versus wavelength |
0.3 dB; typ., 0.05dB. |
Side-mode Suppression ratio |
>= 43 dBc |
Relative intensity noise |
< -135 dB |
Power Meter |
|
Wavelength Range |
850 nm to 1620 nm |
Power Measurement Range |
+10 dBm to -65 dBm |
Power Accuracy |
< +/- 4% |
Power repeatability |
< 0.1 dB |
Sweep Function |
|
Sweep Mode |
Fast Continue Spectrum Sweep |
Resolution |
1 pm / 0.01pm |
Sweep speed |
<10s(10pm,41nm) |
Accuracy |
5 pm |
Power Display |
Peak Power Display |
Center Wavelength Function |
3dB Down Center Wavelength Display |
Reference Function |
Testing Reference Setup |
|
|
Power Adaptor Power |
AC 100 - 240 V ± 10%, 48 - 66 Hz, 100 VA max. |
Display |
PC Interface. |
Storage temperature |
−40°C to +80°C |
Operating temperature |
0°C to +45°C |
Dimensions(W*H*L) |
80mm*166mm*250mm |
Weight |
2kg |